1. How the Pre-Scan measurement is performed in detail
- The Scanner moves to the first measuring position.
- The spectrum analyzer is reset
- Basic settings are transferred to the spectrum analyzer
- The spectrum analyzer is set to max hold mode.
(Max hold means that the highest amplitude at each frequency is kept.) - The spectrum analyzer takes a number of sweeps.
- The Scanner moves to the next measuring position.
- Steps 5 to 6 are repeated until all measuring positions are measured.
- The frequency and amplitude of all peaks are acquired.
(The spectrum analyzer decides what is a peak and what is not.) - The trace is acquired.
- The Scanner moves to its starting position.
2. How the Scan measurement is performed in detail
- The spectrum analyzer is reset.
- The Scanner moves to the first measuring position.
- Basic settings are transferred to the spectrum analyzer.
- The spectrum analyzer zooms in on the first selected frequency.
- The spectrum analyzer is set to max hold mode.
(Max hold means that the highest amplitude at each frequency is kept.) - The spectrum analyzer takes a number of sweeps.
- The frequency and amplitude of the highest peak is acquired.
- Max hold mode is cleared.
- The spectrum analyzer zooms in on the next selected frequency.
- Steps 5 to 9 are repeated until all frequencies are measured.
- The Scanner moves to the next measuring position.
- Steps 5 to 11 are repeated until all measuring positions and frequencies are measured.
- The Scanner moves to its starting position.
3. How the Multi-Scan measurement is performed in detail
Basically, a Multi-Scan measurement is a series of Pre-Scan measurements, one in every measuring position.
From these Pre-Scan measurements field plots similar to Scan measurements, are generated. So when viewing the result from a Multi-Scan measurement any frequency can be chosen to look at in detail.
- The spectrum analyzer is reset
- The Scanner moves to the first measuring position.
- Basic settings are transferred to the spectrum analyzer
- The spectrum analyzer is set to max hold mode.
(Max hold means that the highest amplitude at each frequency is kept.) - The spectrum analyzer takes a number of sweeps.
- The trace data is transferred to the software.
- Max hold mode is cleared.
- The Scanner moves to the next measuring position.
- Steps 4 to 8 are repeated until all measuring positions are measured.
- The Scanner moves to its starting position.
4. How the Immunity-Scan measurement is performed in detail
An Immunity-Scan is, very much lika a Multi-Scan measurement, only we measure immunity and not emission.
- The signal generator (SG) and the Error Detection Device (EDD) is reset
- Basic settings are transferred to the SG and the EDD
- The Scanner moves to the first measuring position.
- The lowest frequency and amplitude is sent to the SG.
- Wait for a while (dwell time) to let the DUT react.
- Measure DUT failure using the EDD and the Detection sequence.
- If the DUT does not fail we increase the amplitude and measure again for DUT failure until the DUT fails or we reach the highest amplitude setting.
- Store frequency and amplitude data.
- Set the next frequency and redo from (5) until we have measured all frequencies.
- The Scanner moves to the next measuring position.
- Steps 4 to 10 are repeated until all measuring positions are measured.
- The Scanner moves to its starting position.