• The spectrum analyzer is reset.
  • The Scanner moves to the first measuring position.
  • Basic settings are transferred to the spectrum analyzer.
  • The spectrum analyzer zooms in on the first selected frequency.
  • The spectrum analyzer is set to max hold mode.
    (Max hold means that the highest amplitude at each frequency is kept.)
  • The spectrum analyzer takes a number of sweeps.
  • The frequency and amplitude of the highest peak is acquired.
  • Max hold mode is cleared.
  • The spectrum analyzer zooms in on the next selected frequency.
  • Steps 5 to 9 are repeated until all frequencies are measured.
  • The Scanner moves to the next measuring position.
  • Steps 5 to 11 are repeated until all measuring positions and frequencies are measured.
  • The Scanner moves to its starting position.