• Proceed from the test above.
  • Settings for the pulse generator:
    • single shot pulse,
    • manual trigger,
    • amplitude TTL = 0 – 2 VPP, and
    • duration = 10 ns.
  • Connect the pulse generator to Ext Arm Input.
  • Set Start Arming to EA on the Analyzer. The Analyzer does not measure.
  • Apply one single pulse to Ext Arm Input.
  • The Analyzer measures once and shows 10 MHz on the display.