- Proceed from the test above.
- Settings for the pulse generator:
- single shot pulse,
- manual trigger,
- amplitude TTL = 0 – 2 VPP, and
- duration = 10 ns.
- Connect the pulse generator to Ext Arm Input.
- Set Start Arming to EA on the Analyzer. The Analyzer does not measure.
- Apply one single pulse to Ext Arm Input.
- The Analyzer measures once and shows 10 MHz on the display.