• Recall Defaults
  • Select Period Single A
  • Set DC, 50 Ω for Input A
  • Connect the rear panel output marked 10 MHz Out to Input A. The Analyzer will display 100 ns
  • Increase Hold-Off Time setting (Settings → Measurement → Hold-Off Time) in steps from 0 s to 120 ns
  • While Hold-Off Time is below 100 ns the result will be about 100 ns (one period). As soon as Hold-Off Time exceeds 100 ns (e.g. 110 ns), the result displayed will be about 200 ns (two periods)