Display modes
Values/Statistics: Numeric display of Measurement values or Statistics parameters with large digits. Values mode also display auxiliary parameter values (with less resolution).
Time-line/Distribution: All measurements are displayed graphically. Multi-channel graphs are colorcoded. Statistics values are also displayed beneath the graphs.
Resolution
Measure up to 4 input signals in parallel with 7 ps resolution per timestamp (Period single, Time Interval, Pulse width, Rise/Fall time, Duty cycle, TIE), or 12 digits/s (frequency and period average).
Smart Frequency/Period avg. calculation mode
Basic resolution for frequency and period average is 12 digits in 1s measuring time. Statistics resolution enhancement algorithm (smart mode) gives up to one extra result digit depending on input signal and measurement setting.
Frequency A, B, D, E
Mode: Parallel measurements on up to 4 inputs. Back-to-back, with or without smart calculation
Range: 0.001 Hz to 400 MHz Aux.
Parameter: Vmax, Vmin, Vp-p
Frequency C (option)
Mode: Back-to-back, with or without smart calculation
Range: See input C
Aux. Parameter: Period C
Frequency Ratio (A,B,C,D, or E) / (A,B,C,D, or E)
Mode: Parallel measurements on 2 or 4 inputs.
Range: (10-9) to 1011
Input Frequency: See input A,B,D,E and C
Aux Parameters: Freq 1, Freq 2
Period A, B, C, D, E average
Mode: Parallel measurements on up to 4 inputs. Back-to-back, with or without smart calculation
Range: See the inverse of Frequency specifications
Aux. Parameter:
- Ch. A, B, D, E: Vmax, Vmin, Vp-p
- Ch. C: Frequency C
TIE A, B, C, D, E (Option 151)
TIE = Time Interval Error, calculated as: Accumulated period - Expected (“ideal”) accumulated period
Mode: Parallel measurements on up to 4 inputs. Back-to-back.
Freq range: See Frequency specifications
Aux. Parameter: Ref Frequency
Period A, B, D or E single
Mode: Parallel measurements on 1 or 2 inputs.
Range 2.5 ns to 1000 sec.
Aux. Parameter (A, B): Vmax, Vmin, Vp-p
Time Interval A, B, D, E (single or continuous)
Mode: Parallel timestamping of trigger events on up to 4 channels on continuous or single-shot signals.
Start and stop channel(s): any of A, B, D, E
Note: each input can produce 1 or 2 trigger events with individual trigger level and slope
Accumulated Time Interval: ON or OFF (adding or subtracting one start channel period to the Time Interval, when required)
Range: -1000s to +1000s
Repetition rate: up to 400 MHz or single-shot events
Min. Pulse width: 1.5 ns
Channel-to-channel offset: 100ps rms typ.
Positive and Negative Pulse Width A, B, D, E
Mode: Parallel measurements on 1 or 2 inputs.
Range: 1.5 ns to 1000 sec.
Repetition rate: up to 300 MHz or single-shot events
Resolution: <10 ps per calculated result
Rise / Fall Time A, B, D, E
Mode: Parallel measurements on 1 or 2 inputs of Rise OR Fall time, or
Mode 2: Single input measurement of Rise AND Fall time on the same pulse.
Range: 1.5 ns to 1000 sec.
Aux. Parameters: Slew rate, Vmax, Vmin
Positive and negative Slew Rate A, B, D, E
Mode: Parallel measurements on 1 or 2 inputs
Calculation: (80% of Vp-p) / (Rise or Fall Time)
Aux. Parameters: Slew rate, Vmax, Vmin
Positive and Negative Duty Cycle A, B, D, E
Mode: Single input meaurement
Range: 0.000001 to 0.999999
Repetition rate: up to 300 MHz
Aux. parameters: Period, Pulse width
Phase A Relative B, B Relative A
Mode: Intended for phase shift or delay measurements of two signals with identical frequency
Accumulated Phase: OFF or ON (adding or subtracting 360o to the Phase, when required)
Range: -180° to +180° (Acc. Phase is OFF)
Resolution: 0.00003° to 100 kHz, decreasing to 0.03° >100 MHz. (10k sample statistics averaging)
Freq. Range: up to 300 MHz
Aux. Parameters: Freq (A), Va/Vb (in dB)
Totalize A, B, D, E
Inputs: up to 4 inputs (A, B, D, E)
Mode: Tot A, B, D, E; Tot A+B, D+E; Tot A-B, D-E: Tot A/B, D/E
Range: 1 to 1010 counts
Range: up to 400 MHz
Start control: Manual, start arming
Stop control: Manual, stop arming, timed
Vmax, Vmin, Vp-p A, B, D, E
Range: -5 V to +5 V, -50V to +50V
Freq. Range: DC, 100Hz to 200 MHz
Coupling: Sine (AC or DC), Square (DC only)
Resolution: 1 mV (5V range), 10 mV (50V range)
Uncertainty (5V range):
- DC, 1Hz to 1kHz: <1% +15 mV
- 1kHz to 20 MHz sine: 3% +15 mV (typ.)
- 20 to 100 MHz sine: 10% +15 mV (typ.)
- 100 to 200 MHz sine: 30% +15 mV (typ.)
(For square waves add 10% to Vmax,/min & 20% to Vp-p) (For 50V range, add 2% + 150 mV)
Aux parameters: Vmin, Vmax, Vp-p