1. Impedance

The input impedance can be set to 1 MΩ or 50 Ω by toggling the corresponding softkey.

CAUTION: Switching the impedance to 50 Ω when the input voltage is above 12 Vrms may cause perma­nent damage to the input circuitry.

2. Attenuation

The input signal’s amplitude can be attenuated by 1 or 10 by toggling the softkey marked 1x/10x. Use attenuation whenever the input signal ex­ceeds the dynamic input voltage range ±5 V or else when attenuation can reduce the influence of noise and interference. See the section deal­ing with these matters at the end of this chap­ter.

3. Coupling

Switch between AC coupling and DC cou­pling by toggling the softkey AC/DC.

Fig. 3-3 AC coupling a symmetrical signal.

Use the AC coupling feature to eliminate un­wanted DC signal components. Always use AC coupling when the AC signal is superim­posed on a DC voltage that is higher than the trigger level setting range. However, we rec­ommend AC coupling in many other measure­ment situations as well.

When you measure symmetrical signals, such as sine and square/triangle waves, AC cou­pling filters out all DC components. This means that a 0 V trigger level is always cen­tered around the middle of the signal where triggering is most stable.

Fig. 3-4 Missing trigger events due to AC coupling of signal with varying duty cycle.

Signals with changing duty cycle or with a very low or high duty cycle do require DC coupling. Fig. 3-4 shows how pulses can be missed, while Fig. 3-5shows that triggering does not occur at all because the signal ampli­tude and the hysteresis band are not centered.

NOTE: For explanation of the hysteresis band, see page 4-3.

Fig. 3-5 No triggering due to AC coupling of signal with low duty cycle.

4. Filter

If you cannot obtain a stable reading, the sig­nal-to-noise ratio (often designated S/N or SNR) might be too low, probably less than 6 to 10 dB. Then you should use a filter. Certain conditions call for special solutions like highpass, bandpass or notch filters, but usu­ally the unwanted noise signals have higher frequency than the signal you are interested in. In that case you can utilize the built-in lowpass filters. There are both analog and dig­ital filters, and they can also work together.

Fig. 3-6 The menu choices after selecting FILTER.

Analog Lowpass Filter

The counter has analog LP filters of RC type, one in each of the channels A and B, with a cutoff frequency of approximately 100 kHz, and a signal rejection of 20 dB at 1 MHz.

Accurate frequency measurements of noisy LF signals (up to 200 kHz) can be made when the noise components have significantly higher frequencies than the fundamental signal.

Digital Lowpass Filter

The digital LP filter utilizes the Hold-Off function described below.

With trigger Hold-Off it is possible to insert a deadtime in the input trigger circuit. This means that the input of the counter ignores all hysteresis band crossings by the input signal during a preset time after the first trigger event.

When you set the Hold-Off time to approx. 75% of the cycle time of the signal, erroneous triggering is inhibited around the point where the input signal returns through the hysteresis band. When the signal reaches the trigger point of the next cycle, the set Hold-Off time has elapsed and a new and correct trigger will be initiated. Instead of letting you calculate a suitable Hold-Off time, the counter will do the job for you by converting the filter cutoff frequency you enter via the value input menu below to an equivalent Hold-Off time.

Fig. 3-7 Value input menu for setting the cutoff frequency of the digital filter.

You should be aware of a few limitations to be able to use the digital filter feature effectively and unambiguously. First you must have a rough idea of the frequency to be measured. A cutoff frequency that is too low might give a perfectly stable reading that is too low. In such a case, triggering occurs only on every 2nd, 3rd or 4th cycle. A cutoff frequency that is too

high (>2 times the input frequency) also leads to a stable reading. Here one noise pulse is counted for each half-cycle.

Use an oscilloscope for verification if you are in doubt about the frequency and waveform of your input signal. The cutoff frequency setting range is very wide: 1 Hz – 50 MHz

Fig. 3-8 Digital LP filter operates in the measuring logic, not in the input amplifier.

5. Man/Auto

Toggle between manual and automatic trigger­ing with this softkey. When Auto is active the counter automatically measures the peak-to-peak levels of the input signal and sets the trigger level to 50% of that value. The attenuation is also set automatically.

At rise/fall time measurements the trigger lev­els are automatically set to 10% and 90% of the peak values.

When Manual is active the trigger level is set in the value input menu designated Trig. See below. The current value can be read on the display before entering the menu.

Speed

The Auto-function measures amplitude and calculates trigger level rapidly, but if you aim at higher measurement speed without having to sacrifice the benefits of automatic trigger­ing, then use the Auto Trig Low Freq func­tion to set the lower frequency limit for volt­age measurement.

If you know that the signal you are interested in always has a frequency higher than a cer­tain value flow , then you can enter this value from a value input menu. The range for flow is 1 Hz to 100 kHz, and the default value is 100 Hz. The higher value, the faster measure­ment speed due to more rapid trigger level voltage detection.

Even faster measurement speed can be reached by setting the trigger levels manually. See Trig below.

Follow the instructions here to change the low-frequency limit:

  • Press SETTINGS->Misc->Auto Trig Low Freq.
  • Use the UP/DOWN arrow keys or the nu­meric input keys to change the low fre­quency limit to be used during the trigger level calculation, (default 100 Hz).
  • Confirm your choice and leave the SET­TINGS menu by pressing EXIT/OK three times.

6. Trig

Value input menu for entering the trigger level manually.

Use the UP/DOWN arrow keys or the nu­meric input keys to set the trigger level. A blinking underscore indicates the cursor po­sition where the next digit will appear. The LEFT arrow key is used for correction, i.e. deleting the position preceding the current cursor position.

Fig. 3-9 Value input menu for setting the trigger level.

NOTE:  It is probably easier to make small ad­justments around a fixed value by us­ing the arrow keys for incrementation or decrementation. Keep the keys de­pressed for faster response

NOTE:   Switching over from AUTO to MAN Trig­ger Level is automatic if you enter a trigger level manually.

Auto Once

Converting “Auto” to “Fixed”

The trigger levels used by the auto trigger can be frozen and turned into fixed trigger levels simply by toggling the MAN/AUTO key. The current calculated trigger level that is visible on the display under Trig will be the new fixed manual level. Subsequent measurements will be considerably faster since the signal levels are no longer monitored by the instrument. You should not use this method if the signal levels are unstable.

NOTE: You can use auto trigger on one input and fixed trigger levels on the other.