1. Configuration parameters for :SYSTEM:CONFIGURE
1.1. TriggerMode
This parameter specifies how the instrument selects trigger levels for measurement comparators. Trigger levels can be set automatically by the instrument or specified by the user.
Parameter syntax: TriggerMode<Input>
where <Input>
is one of A, B, D, E, for example: TriggerModeA
Possible values:
, Auto
, Relative
Manual
Default value (RST condition): Auto
Description
Auto
– Preliminary voltage measurement will be performed to find out signal voltage range and set trigger levels to best values for selected measurement function.
Depending on measurement function selected trigger levels will be automatically set to percentage values of signal voltage range:
- 10% for the main comparator, 90% for the supplementary comparator for signal edge related functions: RiseTime/FallTime/RiseFallTime/PositiveSlewRate/NegativeSlewRate
- 70% for the main comparator, 30% for the supplementary comparator for the functions with hysteresis on: Frequency/FrequencyRatio/frequencySmart/PeriodAverage/PeriodAverageSmart/TIE
- 50% for both main and supplementary comparators for all other functions
Relative
– Preliminary measurement is preformed, but you can adjust trigger levels relative to the measured signal voltage range, in percents, using RelativeTriggerLevel parameter.
Manual
– Trigger levels are set accordingly to values specified by AbsoluteTriggerLevel parameter. It is recommended to use this mode only in cases when Autoset fails to find the best settings.
Example:
SYST:CONF "TriggerModeA=Auto"
1.2. AbsoluteTriggerLevel
Set absolute trigger levels for main and supplementary comparators
Parameter syntax: AbsoluteTriggerLevel<Comparator>
where <Comparator>
is one of A, A2, B, B2, D, D2, E, E2. For example: AbsoluteTriggerLevelA2
Possible values:
Any floating point number representing trigger level in Volts, withing voltage measurement range. Voltage measurement range depends on
and Attenuation
parameters. See description of Preamplifier
and Attenuation
parameters.Preamplifier
Default value (RST condition): 0
Description
Sets trigger level in volts on main comparator (if numeric index is omitted, for example: A, B, D, E) or on supplementary comparator (if numeric index is 2, for example: A2, B2, D2, E2) of the input. Main comparator is used for all measurements performed on input, generating events to internal measurement core whenever signal crosses set trigger level. Supplementary comparator can be selected explicitly for some measurements from input (e.g. Time Interval A, A2), generating independent events to internal measurement core whenever signal crosses set trigger level. Otherwise it is used implicitly for frequency average and period average measurements (to assure wide hysteresis), for Rise & Fall Time and Slew Rate.
Example:
SEND → SYST:CONF "TriggerModeA=Manual; AbsoluteTriggerLevelA=-2.5; AbsoluteTriggerLevelA2=2.5; TriggerModeB=Manual; AbsoluteTriggerLevelB=0; AbsoluteTriggerLevelB2=1.35"
1.3. RelativeTriggerLevel
Sets relative trigger levels in percents relative to signal voltage range, for main and supplementary comparators.
Parameter syntax: RelativeTriggerLevel<Comparator>
where <Comparator>
is one of A, A2, B, B2, D, D2, E, E2. For example: RelativeTriggerLevelB2
Possible values: 0 … 100
Default value (RST condition):
70
for the main comparators
30
for the supplementary comparators
Description
Sets trigger level on main (if comparator index is omitted, e.g. just A) or on supplementary comparator (if comparator index is 2, e.g. A2) of input in percents relative to measured signal voltage range. 0% corresponds to signal min value and 100% corresponds to signal max value. Main comparator is used for all measurements performed on input, generating events to internal measurement core whenever signal crosses set trigger level. Supplementary comparator can be selected explicitly for some measurements generating independent events to internal measurement core whenever signal crosses set trigger level. Otherwise it is used implicitly for frequency average and period average measurements (to assure wide hysteresis), for Rise & Fall Time and Slew Rate.
Example
SEND → SYST:CONF "TriggerModeA=Relative; RelativeTriggerLevelA=65; RelativeTriggerLevelA2=35"
1.4. Slope
Specifies signal slope that triggers comparator
Parameter syntax: Slope<Comparator>
where Comparator
is one of A, A2, B, B2, C, D, D2, E, E2, EA, ER, G, T
Possible values: Positive
, Negative
Default value (RST condition): Positive
Description
Positive
– Positive slope is used for corresponding input
Negative
– Negative slope is used for corresponding input
Example
SYST:CONF "SlopeA=Positive"
1.5. Impedance
Specifies input impedance
Parameter syntax: Impedance<Input>
where <Input>
is one of A, B, D, E. For example: ImpedanceA
Possible values: 50Ohm
, 1MOhm
Default value (RST condition): 1MOhm
Description
50Ohm
– 50 Ohm impedance is used for corresponding input
1MOhm
– 1 MOhm impedance is used for corresponding input
Example
SEND → SYST:CONF "ImpedanceA = 50 Ohm"
1.6. Coupling
Allows to select AC or DC coupling of an input.
Parameter syntax: Coupling<Input>
where <Input>
is one of A, B, D, E. For example: CouplingA
Possible values: DC
, AC
Default value (RST condition): AC
Description
AC coupling allows to block unnecessary DC offset that may be present in the signal.
DC
– DC coupling is used for corresponding input
AC
– AC coupling is used for corresponding input
Example
SEND → SYST:CONF "CouplingA = AC"
1.7. Filter
Selects whether low-pass filter is enabled for an input
Parameter syntax: Filter<Input>
where Input
is one of A, B, D, E. For example: FilterA
Possible values: Off
, 10kHz
, 100kHz
Default value (RST condition): Off
Description
Off
– Filter is not used for corresponding input
10kHz
– 10 kHz low pass filter is used for corresponding input
100kHz
– 100 kHz low pass filter is used for corresponding input
Example
SYST:CONF "FilterA = Off"
1.8. Attenuation
Enables or disables signal attenuation on an input
Parameter syntax: Attenuation<Input>
where Input
is one of A, B, D, E. For example: AttenuationA
Possible values: 1x
, 10x
, Auto
Default value (RST condition): 1x
Description
This setting allow to enable attenuation for situations when measurement signal amplitude is too big for an input.
1x
– 1x attenuation is used for corresponding input (corresponds to no attenuation)
10x
– 10x attenuation is used for corresponding input (signal is decreased 10 times).
Auto
– Attenuation is auto-selected is used for corresponding input.
Correct auto-trigger operation operation and voltage measurements are possible for the following input signal voltage ranges depending on attenuation and preamplifier settings.
Preamplifier=off | Preamplifier=on | |
---|---|---|
Attenuation=1x | -5 … +5 V | -1.5 .. +1.5 V |
Attenuation=10x | -50 … +50 V | -15 .. +15 V |
Attenuation=Auto | -50 … +50 V | -1.5 .. +1.5 V |
Note! When measuring voltage-related signal characteristics, measurement results displayed on the screen and provided by SCPI queries are not decreased 10 times when attenuation is enabled. Even though the signal is attenuated inside the instrument, resulting data are scaled back to original scale.
Example
SYST:CONF "AttenuationA = 10x"
1.9. Preamplifier
Enables or disables signal amplification inside the instrument for an input.
Parameter syntax: Preamplifier<Input>
where Input
is one of A, B, D, E. For example: PreamplifierA
Possible values: Off
, On
Default value (RST condition): Off
Description
Allows to amplify input signal. It is recommended to turn amplification on only if input signal has low amplitude.
Off
– Preamplifier in off for corresponding input
On
– Preamplifier in on for corresponding input
Correct auto-trigger operation and voltage measurements are possible for the following input signal voltage ranges depending on attenuation and preamplifier settings.
Preamplifier=off | Preamplifier=on | |
---|---|---|
Attenuation=1x | -5 … +5 V | -1.5 .. +1.5 V |
Attenuation=10x | -50 … +50 V | -15 .. +15 V |
Attenuation=Auto | -50 … +50 V | -1.5 .. +1.5 V |
Example
SYST:CONF "PreamplifierA = On"
1.10. ArmOn
When arming is enabled, defines whether sample block or each sample within the block should be armed.
Parameter syntax: ArmOn
Possible values: Block
, Sample
Default value (RST condition): Block
Description
When arming is enabled, defines whether sample block or each sample within the block should be armed. Block of samples is a group of samples captured during a measurement (number of samples is defined by SampleCount parameter). This setting is available for modification only when arming is enabled. See StartArmingSource parameter.
Block
– Arm on Block mode: entire sample block is being armed. Please note: when arming on Block and stop arming is not Off – the resulting number of samples in the block might be less than SampleCount.
Sample
– Arm on Sample mode: each individual sample inside the block is being armed.
Example
SYST:CONF "ArmOn = Block"
1.11. Function
Selects measurement function and inputs
Parameter syntax: Function=<function name> <input_or_comparator>[,<input_or_comparator>, ...]
where
<function name>
is one of Frequency, SmartFrequency, PeriodAverage, SmartPeriodAverage, PeriodSingle, TimeInterval, TimeIntervalSingle, AccumulatedTimeInterval, Phase, AccumulatedPhase, TIE, PositiveDutyCycle, NegativeDutyCycle, PositivePulseWidth, NegativePulseWidth, RiseTime, FallTime, RiseFallTime, PositiveSlewRate, NegativeSlewRate, Totalize, TotalizeX+Y, TotalizeX-Y, TotalizeX/Y, Vmin, Vmax, Vpp, Vminmax, DC Offset
input_or_comparator
is one of A, A2, B, B2, C, D, D2, E, E2, EA, ER, G, T. See description below for limitations.
Possible values:
Note, that this parameter is combination of function name, space character and a list of inputs/comparators, separated by comma. Each measurement function supports different maximum number of inputs or comparators. See description below for possible values, their meanings and limitations.
Default value (RST condition): Frequency A
Description
Allows to select measurement function of the instrument.
<function name> | Description |
Frequency | Average frequency over gate time (set by Sample Interval parameter). This is back-to-back measurement (every period of the signal can be captured) for frequencies up to 20 MHz. Up to 4 signals can be measured in parallel. All inputs/comparators can be used. |
FrequencyRatio | Ratio of frequency averages. This mode is just additional math applied over frequency measurements.All inputs/comparators can be used. Minimum two inputs/comparators must be specified. When two inputs/comparators are specified then the frequency ratio of second to first input/comparator is measured. When three inputs/comparators are specified then ratios of 2nd to 1st and 3rd to 1st are measured. When 4 inputs/comparators are specified then ratios of 2nd to 1st and 4th to 3rd are measured. |
SmartFrequency | Smart Frequency function makes use of regression analysis to increase the resolution of the measurement at the expense of measurement speed. Please, note: this mode assumes signal frequency is static within gate time (set by Sample Interval). All inputs/comparators can be used. |
PeriodAverage | Average period over gate time (set by Sample Interval parameter). This is back-to-back measurement (every period of the signal can be captured) for frequencies up to 20 MHz. Up to 4 signals can be measured in parallel. All inputs/comparators can be used. |
SmartPeriodAverage | Smart Period Average function makes use of regression analysis to increase the resolution of the measurement at the expense of measurement speed. Please, note: this mode assumes signal frequency is static within gate time (set by Sample Interval parameter). All inputs/comparators can be used. |
PeriodSingle | Allows to capture single signal periods for periods less than 20 MHz at expense of 50 ns dead-time.All inputs/comparators can be used. Maximum is 2 inputs. |
TimeInterval | Time Interval between up to 4 periodic signals. Result is normalized to the range of -0.5 to +1 signal period.All inputs except C can be used. |
TimeIntervalSingle | Time Interval between single events from up to 4 inputs.All inputs except C can be used. |
AccumulatedTimeInterval | Same as Time Interval, but the result is not normalized.All inputs except C can be used. |
Phase | Phase difference between 2 periodic signals. Result is normalized to the range of -180° to +360°.All inputs except C can be used. |
AccumulatedPhase | Same as Phase, but the result is not normalized.All inputs except C can be used. |
TIE | Time Interval Error (TIE) between up to 4 independent clock sources.All inputs can be used. |
PositiveDutyCycle | Ratio of a pulse signal Positive Pulse Width to its Period.Inputs A, B, D, E can be used. One input only. |
NegativeDutyCycle | Ratio of a pulse signal Negative Pulse Width to its Period.Inputs A, B, D, E can be used. One input only. |
PositivePulseWidth | Positive Pulse Width of a pulse signal.Inputs A, B, D, E can be used. Maximum is 2 inputs. |
NegativePulseWidth | Negative Pulse Width of a pulse signal.Inputs A, B, D, E can be used. Maximum is 2 inputs. |
RiseTime | Measures how much time it takes for the signal to go from 10% to 90% of its voltage range.Inputs A, B, D, E can be used. Maximum is 2 inputs. |
FallTime | Measures how much time it takes for the signal to go from 90% to 10% of its voltage range.Inputs A, B, D, E can be used. Maximum is 2 inputs. |
RiseFallTime | Measures how much time it takes for the signal to go from 10% to 90% of its voltage range and back.Inputs A, B, D, E can be used. One input only. |
PositiveSlewRate | Measures how fast signal voltage increases from 10% to 90% of its range.Inputs A, B, D, E can be used. Maximum is 2 inputs. |
NegativeSlewRate | Measures how fast signal voltage decreases from 90% to 10% of its range.Inputs A, B, D, E can be used. Maximum is 2 inputs. |
Totalize | Counts number of events on up to 4 input channels in parallel.All inputs except C can be used. |
TotalizeX+Y | Totalize with additional maths applied.All inputs except C can be used. Minimum is 2 inputs. |
TotalizeX-Y | Totalize with additional maths applied.All inputs except C can be used. Minimum is 2 inputs. |
TotalizeX/Y | Totalize with additional maths applied.All inputs except C can be used. Minimum is 2 inputs. |
Vmin | Minimum voltage level of a signal.Inputs A, B, D, E can be used. |
Vmax | Maximum voltage level of a signal.Inputs A, B, D, E can be used. |
Vpp | Signal maximum and minimum voltage levels difference.Inputs A, B, D, E can be used. |
Vminmax | Minimum and maximum voltage levels of a signal.Inputs A, B, D, E can be used. One input only. |
DC Offset | Measure DC offset voltage of a signal. Note: DC coupling must be enabled for corresponding input. |
Example
:SYST:CONF "Function = Frequency A"
:SYST:CONF "Function=Period Average A,B2,EA"
1.12. HoldOff
Dead time between consecutive trigger events
Parameter syntax: HoldOff
Possible values: 0 - 2.683 s
Default value (RST condition): 0 s
Description
Adds dead time between consecutive trigger events. Used to cope with contact bouncing or signal oscillations.
Example
SYST:CONF "HoldOff = 0.555 s"
1.13. LimitBehaviour
Enables of disables limits for measurements. Defines instrument behavior when specified limits are exceeded.
Parameter syntax: LimitBehaviour
Possible values: Off
, Capture
, Alarm
, AlarmStop
Default value (RST condition): Off
Description
Defines how the instrument will react on limits. Limit criterion is set by Limit Type, Upper Limit and Lower Limit. For all limit behavior choices except Off the following is true: If measured value has fell off the limit criterion during measurement session then red exclamation mark indicator is displayed.
Off
– Limits are disabled.
Capture
– Only samples meeting the limit criterion are captured, the rest are discarded. Limit status is displayed
Alarm
– All samples are captured, limit status is displayed
AlarmStop
– Measurement session stops if measured value doesn’t meet the limit criterion
Example
SYST:CONF "LimitBehaviour = Alarm"
1.14. LimitLower
Specifies lower limit.
Parameter syntax: LimitLower
Possible values: any decimal
Default value (RST condition): 0
Description
Lower limit (used if LimitType=Above
or LimitType=Range
and LimitBehaviour
is not Off
)
Example
SYST:CONF "LimitLower = 0 Hz"
1.15. LimitSeriesName
Specifies name of series for which limit is applies to.
Parameter syntax: LimitSeriesName
Possible values: All
, A
,B
,A/B
,A+B
,Vmin
, etc.
Default value (RST condition): All
Description
Name of the series that the limit is applied to, or “All“ if applied to all series. This parameter can be used only when limits are enabled (LimitBehaviour
is not Off
)
Example
SYST:CONF "LimitSeriesName = A/B"
1.16. LimitType
Specifies limit type.
Parameter syntax: LimitType
Possible values: Above
, Below
, Range
Default value (RST condition): Above
Description
Above
– Results above set Lower Limit will pass
Below
– Results below set Upper Limit will pass
Range
– Results within the set limits will pass
This parameter can be used only when limits are enabled (LimitBehaviour
is not Off
)
Example
SYST:CONF "LimitType = Range"
1.17. LimitUpper
Specifies upper limit
Parameter syntax: LimitUpper
Possible values: any decimal
Default value (RST condition): 0
Description
Upper limit (used if LimitType=Below
or LimitType=Range
and LimitBehaviour
is not Off
)
Example
SYST:CONF "LimitUpper = 24.7 Hz"
1.18. MathCoeffK, MathCoeffL, MathCoeffM
Specifies values for coefficients K, L, M, when mathematical formula is enabled.
Parameter syntax: MathCoeff<Coefficient>
where Coefficient
is one of K, L, M,. For exampe: MathCoeffL
Default value (RST condition):
MathCoeffK = 1
MathCoeffL = 0
MathCoeffM = 1
Description
K, L, M constants used in Math formula.
Example
SYST:CONF "MathCoeffM = 1"
1.19. MathCustomUnit
Overrides result units when mathematical formula is applied.
Parameter syntax: MathCustomUnit
Description
When math formula is applied, measurement results units are usually determined automatically. This parameter allows to override the unit of the value after math formula is applied. The length of the custom unit must not exceed 4 characters
Example
SYST:CONF "MathCustomUnit = Emu"
1.20. MathMode
Selects mathematical formula to apply for results
Parameter syntax: MathMode
Possible values: Off
, K*X+L
, K/X+L
, (K*X+L)/M
, (K/X+L)/M
, X/M-1
Default value (RST condition): Off
Description
Allows to apply math over measurement results. Please note, some of available formulae change unit of original value.
Example
SYST:CONF "MathMode = K/X+L"
1.21. MathSeriesName
Select series to apply math formula to.
Parameter syntax: MathSeriesName
Possible values: All
, A
, B
, A/B
, A+B
, Vmin
, etc.
Default value (RST condition): All
Description
Name of the series that MathMode is applied to, or “All“ if applied to all series
Example
SYST:CONF "MathSeriesName = A+B"
1.22. PulseOutputMode
Controls the rear Pulse Output of the device.
Parameter syntax: PulseOutputMode
Possible values: Off
, PulseGenerator
, GateOpen
, AlarmOutActiveHigh
, AlarmOutActiveLow
Default value (RST condition): Off
Description
Controls the rear Pulse Output of the instrument. The amplitude is set to TTL levels into 50 Ohm irrespective to output mode.
Off
– No signal on Pulse Output
PulseGenerator
– Continuous pulse train with period and pulse width set in next menus
GateOpen
– Indicates when measurement is in progress
AlarmOutActiveHigh
– High level output when limits alarm is active, low level otherwise
AlarmOutActiveLow
– Low level output when limits alarm is active, high level otherwise
Example
SYST:CONF "PulseOutputMode = AlarmOutActiveLow"
1.23. PulseOutputPeriod
Set period of pulses generated on rear Pulse Output of the instrument.
Parameter syntax: PulseOutputPeriod
Possible values: 10 ns .. 2.147 s
Default value (RST condition): 1 ms
Description
Set period of pulses generated on Pulse Output if Mode is set to Pulse Generator. Resolution for this parameter is 2 ns.
Example
SYST:CONF "PulseOutputPeriod = 12 ns"
1.24. PulseOutputWidth
Set width of pulses generated on rear Pulse Output of the instrument.
Parameter syntax: PulseOutputWidth
Possible values: 4 ns .. 2.146999994 s
(but at least 6 ns lower than PulseOutputPeriod
)
Default value (RST condition): 500 us
Description
Set width of pulses generated on Pulse Output if Mode is set to Pulse Generator. Resolution for this parameter is 2 ns. Should be at least 6 ns less than PulseOutputPeriod.
Example
SYST:CONF "PulseOutputWidth = 2.1 s"
1.25. SampleCount
Defines number of samples to be collected for each measurement series.
Parameter syntax: SampleCount
Possible values: up to 31999999 samples (depends on selected function and inputs/comparators and other settings)
Default value (RST condition): 1
Description
Defines number of samples to be collected for each measurement series.
Example
SYST:CONF "SampleCount = 10000"
1.26. SampleInterval
Specifies how often measurement samples are generated and/or define gate time.
Parameter syntax: SampleInterval
Default value (RST condition): 10 ms
Description
Defines gate time for Frequency measurement. Please note, that if signal period is larger than this value then actual sample interval will be equal to signal period.
For Frequency, Sample Interval should be in the range: 1 us .. 10.995 ks Please note: to allow sample interval below 1 us corresponding license is required.
Example
SYST:CONF "SampleInterval = 10 ms"
1.27. SignalSource
Specifies whether to use input signals for measurement or internal test generator.
Parameter syntax: SignalSource
Possible values: Inputs
, Test
Default value (RST condition): Inputs
Description
Built-in test signal generator is used for performing internal device calibrations but can also be used for testing purposes. This setting must be used only for testing or demonstrational purposes when there is no possibility to connect external source of signal to instrument’s inputs.
Inputs
– Normal operation
Test
– Using internal test signal generator instead of inputs on front (back) panel. Output of the internal generator are connected to inputs A, B, D, E and can be measured. Please, note: test generator is using independent coarse time base and is not expected to provide accurate frequency. It is not the same generator which drives rear Pulse Output of the device.
Example
SYST:CONF "SignalSource = Test"
1.28. StartArmingDelay
Delay for making measurement after arming event.
Parameter syntax: StartArmingDelay
Default value (RST condition): 0 s
Description
Defines time after start arming event when measurement should be started. Start Arming Delay should be in the range: 0 s .. 10.995 ks
Example
SYST:CONF "StartArmingDelay = 8.556 ks"
1.29. StartArmingSlope
Signal slope to use as arming event.
Parameter syntax: StartArmingSlope
Possible values: Positive
, Negative
Default value (RST condition): Positive
Description
Slope which arms measurement
Negative
– Negative slope is used for Start Arming
Positive
– Positive slope is used for Start Arming
Example
SYST:CONF "StartArmingSlope = Negative"
1.30. StartArmingSource
Input or comparator that will be used to detect arming event.
Parameter syntax: StartArmingSource
Possible values: Off
, EA
, A
, B
, D
, E
, A2
, B2
, D2
, E2
Default value (RST condition): Off
Description
Defines whether signal from one of device inputs should be used to arm start of measurement.
Example
SYST:CONF "StartArmingSource = A"
1.31. StopArmingDelay
Delay for stopping measurement after stop arming event.
Parameter syntax: StopArmingDelay
Default value (RST condition): 0 s
Description
Defines time after stop arming event when measurement should be stopped. Stop Arming Delay should be in the range: 0 s .. 10.995 ks
Example
SYST:CONF "StopArmingDelay = 6.652 ks"
1.32. StopArmingSlope
Signal slope to use as stop arming event.
Parameter syntax: StopArmingSlope
Possible values: Positive
, Negative
Default value (RST condition): Positive
Description
Slope which arms measurement stop
Negative
– Negative slope is used for Stop Arming
Positive
– Positive slope is used for Stop Arming
Example
SYST:CONF "StopArmingSlope = Negative"
1.33. StopArmingSource
Input or comparator to use for stop arming event detection.
Parameter syntax: StopArmingSource
Possible values: Off
, EA
, A
, B
, D
, E
, A2
, B2
, D2
, E2
Default value (RST condition): Off
Description
Defines whether signal from one of device inputs or timer should be used to stop measurement.
Example
SYST:CONF "StopArmingSource = B2"
1.34. TestSignalFrequency
Frequency of internal built-in test generator.
Parameter syntax: TestSignalFrequency
Possible values: 1.039 kHz .. 68 MHz
Default value (RST condition): 1 MHz
Description
Sets frequency of built-in test generator. Please, note: test generator is using independent coarse timebase and is not expected to provide accurate frequency. It is not the same generator which drives rear Pulse Output of the device. Test Signal Frequency should be in the range: 1.039 kHz .. 68 MHz
Example
SYST:CONF "TestSignalFrequency = 5.555 kHz"
1.35. TieReferenceFrequency A, A2, B, B2, D, D2, E, E2, EA, ER
Reference frequency for input or comparator when doing TIE measurements.
Parameter syntax: TieReferenceFrequency<Input>
where Input
is one of A
, A2
, B
, B2
, D
, D2
. For example: TieReferenceFrequencyA
Possible values: 100 mHz .. 400 MHz
Default value (RST condition): 10 MHz
Description
Reference frequency for input A, A2, B, B2, D, D2, E, E2, EA, ER. Should be in the range: 100 mHz .. 400 MHz
Example
SYST:CONF "TieReferenceFrequencyB = 101 mHz"
1.36. TieReferenceFrequencyC
Reference frequency for input C when doing TIE measurements.
Parameter syntax: TieReferenceFrequencyC
Possible values: 100 mHz .. 24 GHz
Default value (RST condition): 1 GHz
Description
Reference frequency for input C. Ref Frequency for input C should be in the range: 100 mHz .. 24 GHz
Example
SYST:CONF "TieReferenceFrequencyC = 12 GHz"
1.37. TieReferenceFrequencyDetection
Defines if reference frequency is detected automatically or should be set manually
Parameter syntax: TieReferenceFrequencyDetection
Possible values: On
, Off
Default value (RST condition): On
Description
Defines if reference frequency is detected automatically or should be set manually
On
– Reference frequency for TIE measurement is detected automatically
Off
– Reference frequency for TIE measurement is set manually
Example
SYST:CONF "TieReferenceFrequencyDetection = Off"
1.38. TieReferenceFrequencyNumberOfDigits
Number of digits detected reference frequency should be rounded to.
Parameter syntax: TieReferenceFrequencyNumberOfDigits
Possible values: 0 .. 10
Default value (RST condition): 5
Description
Defines how many digits detected reference frequency should be rounded to. Ref Frequency Number Of Digits should be in the range: 0 .. 10
Example
SYST:CONF "TieReferenceFrequencyNumberOfDigits = 9"
1.39. TimebaseReference
Specifies which reference clock to use for measurements.
Parameter syntax: TimebaseReference
Possible values: Auto
, Internal
, External
Default value (RST condition): Auto
Description
Defines which reference clock will be used for measurement.
Internal
– Internal timebase reference is used
External
– External timebase reference is used if it is connected and within expected parameters, otherwise measurement is not performed.
Auto
– External timebase reference is used if it is connected and within expected parameters, otherwise internal
Example
SYST:CONF "TimebaseReference = External"
1.40. Timeout
Enable of disable measurement timeout.
Parameter syntax: Timeout
Possible values: On
, Off
Default value (RST condition): Off
Description
Used to make measurement session to end if signal is missing for more than specified Timeout Time.
On
– Timeout is on
Off
– Timeout is off
Example
SYST:CONF "Timeout = Off"
1.41. TimeoutTime
Set measurement timeout value.
Parameter syntax: TimeoutTime
Possible values: 10 ms .. 1 ks
Default value (RST condition): 100 ms
Description
If Timeout is On, measurement session ends if signal is missing for more than specified Timeout Time.
Example
SYST:CONF "TimeoutTime = 168 ms"
1.42. VoltageMode
Parameter syntax: VoltageMode
Possible values: Normal
, VerySlow
, Slow
, Fast
, VeryFast
Default value (RST condition): Normal
Description
Defines minimal signal frequency for which voltage measurements and/or auto-trigger works correctly. Please, note: voltage/auto-trigger for lower frequencies are measured at the expense of measurement speed. It is recommended to treat this setting as a fallback for the cases where Autoset fails to find best setting automatically.
VerySlow
– Used for signals with frequency in range: < 10 Hz
Slow
– Used for signals with frequency in range: 10 Hz to 100 Hz
Normal
– Used for signals with frequency in range: 100 Hz to 1 kHz
Fast
– Used for signals with frequency in range: 1 kHz to 10 kHz
VeryFast
– Used for signals with frequency 10 kHz and above.
DC signals can be measured with any mode.
1.43. InternalCalibrationMode
Defines when to start internal calibration.
Parameter syntax: InternalCalibrationMode
Possible values: Every30Min
, BeforeEveryMeasurement
, OnceAfterWarmup
Default value (RST condition): Every30Min
Description
Internal calibration is always done on device start-up and after it has warmed up. More frequent calibration can be done in order to improve timing measurement resolution.
Every30Min
– Additional calibration every 30 minutes
BeforeEveryMeasurement
– Additional calibration before every measurement
OnceAfterWarmup
– No additional calibration
Example
SYST:CONF "InternalCalibrationMode = Every30Min"
1.44. NumOfBlankDigits
Number of digits to blank on display.
Parameter syntax: NumOfBlankDigits
Possible values: 0 .. 15
Default value (RST condition): 0
Description
Defines the number of least significant digits to be masked. This can be used to help operator to read out the results of jittery measurements. Please note: it applies only to the current measured value, not statistics. Digits Blank should be in the range: 0 .. 15
Example
SYST:CONF "NumOfBlankDigits = 12"
1.45. ScreenSaverTimeout
Timeout to switch off display.
Parameter syntax: ScreenSaverTimeout
Possible values: 5minutes
, 10minutes
, 30minutes
, 1hour
, Never
Default value (RST condition): 10minutes
Description
Inactivity timeout for turning device display off
5minutes
– Inactivity timeout set to 5 minutes
10minutes
– Inactivity timeout set to 10 minutes
30minutes
– Inactivity timeout set to 30 minutes
1hour
– Inactivity timeout set to 1 hour
Never
– Device display is always on
Example
SYST:CONF "ScreenSaverTimeout = Never"
1.46. Brightness
Display brightness.
Parameter syntax: Brightness
Possible values: Minimum
, Low
, Medium
, High
, Maximum
Default value (RST condition): Maximum
Description
Inactivity timeout for turning device display off
Minimum
– Minimum brightness
Low
– Low brightness
Medium
– Medium brightness
High
– High brightness
Maximum
– Maximum brightness
Example
SYST:CONF "Brightness = Minimum"
1.47. Wired and Wireless network settings
IPAddress/WirelessIPAddress
Specifies static (manual) IPv4 address in Ethernet network
Parameter syntax: IPAddress
, WirelessIPAddress
Possible values: IP address (four numbers, dot-separated)
Default value (RST condition): 192.168.0.99
Description
IP address for Ethernet interface when static IP address mode is enabled in IPMode
parameter. When DHCP mode is enabled, the IP is assigned by a DHCP server in network and can be read back with SYST:CONF? query)
IP address for Wireless interface (with DHCP, configured address can be read back)
Example
SYST:CONF "IPAddress = 192.168.0.99"
SYST:CONF "WirelessIPAddress = 192.168.0.99"
IPDNS/WirelessIPDNS
DNS servers
Parameter syntax: IPDNS1
, WirelessIPDNS1
and IPDNS2
, WirelessIPDNS2
Possible values: IP address (four numbers, dot-separated)
Default value (RST condition): 8.8.8.8
Description
DNS server(s) IP addresses for Ethernet interface when static IP address mode is enabled in IPMode
parameter. When DHCP mode is enabled, these settings are assigned by a DHCP server in network and can be read back with SYST:CONF? query)
1st nameserver IP address for Wireless interface (with DHCP, configured address can be read back)
Example
SYST:CONF "IPDNS1 = 8.8.8.8; IPDNS2 = 1.1.1.1"
SYST:CONF "WirelessIPDNS1 = 8.8.8.8; WirelessIPDNS2 = 1.1.1.1"
IPGateway/WirelessIPGateway
Default gateway.
Parameter syntax: IPGateway
, WirelessIPGateway
Possible values: IP address (four numbers, dot-separated)
Default value (RST condition): 192.168.0.1
Description
Gateway IP for Ethernet interface when static IP address mode is enabled in IPMode
parameter. When DHCP mode is enabled, this setting is assigned by a DHCP server in network and can be read back with SYST:CONF? query)
Gateway IP address for Wireless interface (with DHCP, configured address can be read back)
Example
SYST:CONF "IPGateway = 192.168.0.1"
SYST:CONF "WirelessIPGateway = 192.168.0.1"
IPMode/WirelessIPNetmask
Select whether to configure IP address automatically from DHCP server or manually
Parameter syntax: IPMode
, WirelessIPMode
Possible values: DHCP
, Static
Default value (RST condition): DHCP
Description
IP configuration mode for Ethernet interface
IP netmask for Wireless interface (with DHCP, configured netmask can be read back)
DHCP
– IP settings are aquired via DHCP
Static
– Static IP configuration. When this mode is selected you should also configure IPAddress and IPNetmask, IPGateway, IPDNS1 (and optionally IPDNS2) parameters.
Example
SYST:CONF "IPMode = DHCP"
SYST:CONF "WirelessIPMode = DHCP"
IPNetmask/WirelessIPNetmask
Parameter syntax: IPNetmask
, WirelessIPNetmask
Possible values: IP mask (four numbers, dot-separated)
Default value (RST condition): 255.255.255.0
Description
IP netmask for Ethernet interface when static IP address mode is enabled in IPMode
parameter. When DHCP mode is enabled, this setting is assigned by a DHCP server in network and can be read back with SYST:CONF? query)
IP netmask for Wireless interface (with DHCP, configured netmask can be read back)
Example
SYST:CONF "IPNetmask = 255.255.255.0"
SYST:CONF "WirelessIPNetmask = 255.255.255.0"