Selects whether low-pass filter is enabled for an input
Parameter syntax: Filter<Input>
where Input is one of A, B, D, E. For example: FilterA
Possible values: Off, 10kHz, 100kHz
Default value (RST condition): Off
Description
Off - Filter is not used for corresponding input
10kHz - 10 kHz low pass filter is used for corresponding input
100kHz - 100 kHz low pass filter is used for corresponding input
Example
SYST:CONF "FilterA = Off"