Controls the rear Pulse Output of the device.
Parameter syntax: PulseOutputMode
Possible values: Off, PulseGenerator, GateOpen, AlarmOutActiveHigh, AlarmOutActiveLow
Default value (RST condition): Off
Description
Controls the rear Pulse Output of the instrument. The amplitude is set to TTL levels into 50 Ohm irrespective to output mode.
Off - No signal on Pulse Output
PulseGenerator - Continuous pulse train with period and pulse width set in next menus
GateOpen - Indicates when measurement is in progress
AlarmOutActiveHigh - High level output when limits alarm is active, low level otherwise
AlarmOutActiveLow - Low level output when limits alarm is active, high level otherwise
Example
SYST:CONF "PulseOutputMode = AlarmOutActiveLow"