Controls the rear Pulse Output of the device.
Parameter syntax: PulseOutputMode
Possible values: Off
, PulseGenerator
, GateOpen
, AlarmOutActiveHigh
, AlarmOutActiveLow
Default value (RST condition): Off
Description
Controls the rear Pulse Output of the instrument. The amplitude is set to TTL levels into 50 Ohm irrespective to output mode.
Off
- No signal on Pulse Output
PulseGenerator
- Continuous pulse train with period and pulse width set in next menus
GateOpen
- Indicates when measurement is in progress
AlarmOutActiveHigh
- High level output when limits alarm is active, low level otherwise
AlarmOutActiveLow
- Low level output when limits alarm is active, high level otherwise
Example