Reference frequency for input or comparator when doing TIE measurements.

Parameter syntax: TieReferenceFrequency<Input>

where Input is one of A, A2, B, B2, D, D2. For example: TieReferenceFrequencyA

Possible values: 100 mHz .. 400 MHz

Default value (RST condition): 10 MHz

Description

Reference frequency for input A, A2, B, B2, D, D2, E, E2, EA, ER. Should be in the range: 100 mHz .. 400 MHz

Example

SYST:CONF "TieReferenceFrequencyB = 101 mHz"